INSTALLATION • OPERATION • MAINTENANCE
Conductivity/TDS Models: 756II, 757II, 758II & 759II
Resistivity Models: 751II, 752II, 753II & 754II
750 Series II
CONDUCTIVITY/TDS
&
RESISTIVITY
MONITOR/CONTROLLER
Operation
Manual
09 August 11
ACCURACY • RELIABILITY • SIMPLICITY
Water Quality Instrumentation
Accuracy Reliability Simplicity
Vue de la page 0
1 2 3 4 5 6 ... 75 76

Résumé du contenu

Page 1 - 09 August 11

INSTALLATION • OPERATION • MAINTENANCEConductivity/TDS Models: 756II, 757II, 758II & 759IIResistivity Models: 751II, 752II, 753II & 754II750

Page 2

NOTE: 3 1/2 digit LCD may be replaced with either a 3 1/2 or a 4 1/2 digit backlit LCD - available as options. ACCURACYAnalog Models: + 2% of Full Sca

Page 3 - QUICK REFERENCE GUIDE

E. SENSORS 1. CONDUCTIVITY/ TDS 750 Series II Conductivity/TDS Monitor/controllers use the CS51LC, CS51 or CS52 Series sensor depending on range.

Page 4 - Section V.A

RANGEResistivity0-20MΩ0-10MΩ†0-5MΩ†0-2MΩ0-1MΩ†0-500KΩ†0-200KΩConductivity0-1 µS*†0-1 ppm*†0-2 µS*†0-2 ppm*†0-5 µS*†0-5 ppm*†0-10 µS†0-10 ppm†0-

Page 5 - 750 Series II

A. GENERALThis section provides the recommended procedures for properly installing the 750 Series II Conductivity/Resistivity Monitor/controller, a

Page 6 - TABLE OF CONTENTS

4. Slide the enclosure through the panel cutout   untilitsangecontactsthepanel. 5. Insertmountingscrewsthroughtheange mounting

Page 7 - TABLE OF CONTENTS Continued

D. SENSOR INSERTION/IMMERSION MOUNTING The Sensor’s mounting orientation must provide a continuous andadequatecirculationowtopreventthetrap

Page 8

4. Turn the front panel around so that the back side is facing you and set aside for now. 5. Carefully remove front panel, leaving the harness

Page 9 - I. INTRODUCTION

  d. Install5.49kΩresistoratRED(RD)andthe NEUTRAL (NU) connector locations, as shown in    gureII.E.5. 5. Carefullyreinstal

Page 10 - DOUBLE INSULATED

14 c. Connect to your own alarm or ?Use the following as guidelines. Connector is a standard 2 wire Methode* style connector. Connecto

Page 11 - EXAMPLE:

F. 0-10 VDC RECORDER OUTPUTThe0-10VDCoutputisdesignedtogivetheuserthecapabilityof sending a signal to a remote meter, recorder, PLC or SC

Page 13 - II. INSTALLATION

16G. RE-RANGE YOUR MONITOR/CONTROLLER (Range Module Installation) 1. DESCRIPTIONThe 750 Series II Monitor/controllers have been design

Page 14

17 3. CHANGING ANALOG METER SCALE (DIAL)After changing ranges on an analog model 750II it may be nec-essary to change the meter scale (dial). The

Page 15 - 3/4" FNPT

18 4. LABEL CHANGESince you have re-ranged your Monitor/controller it may be necessary to relabel it. If you made any one of the following changes

Page 16 - Figure II.E.2

19A. SC/SCO MODULE (SECOND ALARM/CONTROL OPTION) (Digital & OEM Models ONLY)-SC Second Alarm Control Module ordered with

Page 17

20Front PanelWARNING: There are two (2) O-rings installed on the switches, one (1) on the shank and the other is under the push button. Both of these

Page 18 - PANEL MOUNTED DISPLAY

The following instructions describe the steps for converting the Monitor/controller.RefertogureIII.A.4forthelocationoftheSPC(s) on your Mon

Page 19 - Main CB Assembly

c. Hysteresis (Dead Band) Adjustment*Thehysteresisordeadbandisapproximately±3%ofthesetpoint at full scale as it leaves the factory. Unde

Page 20 - REMOVE TO INSTALL

B. 4A/4AO MODULE (4-20mA OPTION)Available only on the digital models 758II, 759II, 753II and 754II, or all analog and digital OEM Monitor/controlle

Page 21 - BROWN WIRERED WIRE

24FATAL TO PERSONNEL. ONLY QUALIFIED PERSON-NEL SHOULD INSTALL ELECTRICAL EQUIPMENT.PhysicalIf the front panel has all ready been removed from

Page 22 - FRONT PANEL

25 5. If not, adjust the CAL control marked “4mA” until   theDVMindicates4mA,seegureIII.B.3. 6. PresstheFullScaleTestSwitch,theD

Page 23 - ACCESSORIES

If you read nothing else in this manual please read this Quick Reference Guide. PLEASE READ and COMPREHEND ALL WARNINGS, CAUTIONS and ADVISEMENTS CONT

Page 24 - Figure III.A.3

4. CONVERTING A CURRENT TO A VOLTAGECurrent measuring devices actually measure voltage, but have aninternalresistorasshowni

Page 25 - CM NO NC

27C. TEMPERATURE MODULE (Digital Models only)-TP Temperature Module ordered with Monitor/controller.TPO Temperature Module Kit ordere

Page 26 - MICROSIEMENS / CM

28 5. Set front panel down or carefully allow to hang from the harness. Do not drop as the harness connector will pull out allowing the fr

Page 27

29 a. TPC “Calibration” Module ProcedureThis procedure assumes the front panel is removed. 1. Ensure power is OFF. 2. Remove the sensor lead

Page 28 - 4A/4AO - 4-20mA MODULE

30 19. Continue or reinstall the front panel and tightly secure four (4) retaining screws, see REASSEMBLY below.*CAUTION:Thesensorinputand0

Page 29 - Figure III.B.5

31 counterclockwise position. If not, turn SCO Set Point control fully CCW - it may be 30 turns - or until it clicks.   Forlocationseeg

Page 30

32 5. OEM INSTALLATIONa. Single Display ModelsAn OEM installation requires an additional display harness, part number TPMH. This harness is insta

Page 31 - BR BL OR - +

33 6. OEM INSTALLATIONb. Dual Display Models SPECIAL ORDER These models allow simultaneous display of both Cond/TDS, RES or pH/ORP and Temperatur

Page 32 - WARNING:

34D. ALARM/CONTROL KIT (Digital & OEM Models only) -TH Controller wiring Harness (requires second alarm/control module) orde

Page 33 - FRONT PANEL DISPLAY

by pulling lightly on the wires. It may be necessary    touseasmallatscrewdrivertoslightlyspreadthe    retainingring.Seegu

Page 34

connected to an appropriate Sensor, and that the Sensor is immersed in water within the range that the Monitor/controller will be required to read; an

Page 35 - EASSEMBLY

c. Remove harness from SCO Module by pulling on    harness,seegureIII.D.3. d. Set SCH harness aside, it is no longer used. 4. Re

Page 36

8. Turn SCO Set Point control fully counter clockwise   —itmaybe30turns-untilitclicks.SeegureIII.D.7 for location. 9. Locate

Page 37

38Set Point Conversion (SPC) / Reversing Set PointTheMyronLCompanyshipstheTemperaturecontrollercong-ured to Alarm/control when the temperatur

Page 38 - TP/TPO - TEMPERATURE MODULE

39E. INSTALLATION OF BOTH TEMPERATURE MODULE & 4-20mA OUTPUT MODULE (Digital and OEM Models ONLY) -TP/TPO Temperature Modu

Page 39 - BR BL OR - +

40F. 3 SENSOR INPUT MODULE3SO 3 Sensor Input Module with 3 foot / 1 meter cable ordered separately (customer installed and mounted).3SE

Page 40

Module to the main Monitor/controller CB at the sensor input location as shown on label. 5. Attach the separate small cable harness from

Page 41

42  gureIII.F.4.To disable the Sensor alarm/control function completely. 1. Locate the 8 position select switch on the 3S Module. 2. Set all

Page 42

43G. 3 RANGE SWITCH MODULE 3RO 3 Range Switch Module with 3 foot / 1 meter cable ordered separately (customer installed and mounted).3RE

Page 43 - LEFT SIDE VIEW

443ROModuleasanaccessoryyoumayneedtorecongurethe select switch for it to Range properly. 1. Locate the 8 position select switch on the

Page 44

45H. 3 SENSOR / 3 RANGE MODULE 3SRO 3 Sensor / 3 Range Switch Module with 3 foot / 1 meter cable ordered separately (customer installed and mou

Page 45 - BK WT RD GN NU R- R+

1750 Series IIModel 758 -121-SC(A Digital ConductivityMonitor/controller,with a Range of 0-2000 µS,and a Second Alarm/control,Displaying Full Scale

Page 46 - ENCLOSURE

46Sensor ConnectionsFor convenience, it is recommended you connect the Primary or controlling sensor to SENSOR 1 position as shown in gureIII.H.1.

Page 47

47For Conductivity/TDS - X1, X10 and X100. For Resistivity - X1, 0.1, and 0.01. If adding an alarm/control range. 1. Locate the 8 position select s

Page 48 - BK WT RD GN NU R- R+

48I. PIEZO ELECTRIC ALARM -PA Piezo Electric Alarm ordered with Digital Monitor/controller.PAO Piezo Electric Alarm ordered separate

Page 49

49 a. For a Conductivity/TDS Monitor/controller, as shipped from the Myron L Company, the alarm will sound off when the Full Scale Te

Page 50

50J. REMOTE ALARM - RA™ RA Ordered as an accessory, includes 8” harness with connector, and two (2) wire nuts. 1. DESCRIPTIONThisremote

Page 51

51 alarm for approximately three minutes or until you improve the water quality (readjust controller set point).   Ifthreeminutesmuting

Page 52

52IV. OPERATING PROCEDURESA. FRONT PANEL INDICATORS & CONTROLSThe front panel illustrations to the right, switch and indicator operational desc

Page 53

53B. OEM FRONT PANEL INDICATORS & CONTROLSThe simulated OEM front panel illustrations to the right, and switch and indicator

Page 54 - CONTROLLER

54C. SETUP PROCEDURES These Setup procedures cover (1) converting the alarm circuit to trigger on a decreasing (conductivity) or increasing (Resist

Page 55

55CAUTION: adjusting the hysteresis too tight may cause the alarmtouctuate(on-off)duetoow,chemicalmixingorbub-bles causing the relay to ch

Page 56 - IV. OPERATING PROCEDURES

TABLE OF CONTENTS 750 Series II ILLUSTRATION (758II-121-SC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 57 - TEMPERATURE

56A. PRIMARY COMPONENT IDENTIFICATIONAsidentiedinSectionIII,theMonitors/controllersswitch(es)and indicator components are mounted directl

Page 58 - RESISTIVITY

57B. METER MECHANICAL ZERO PROCEDURESANALOG METERS ONLY 1. Ensure power is OFF. 2. Note the position of the meter needle. If it has come to

Page 59

58 3. Slowly shake the sensor to remove air bubbles from inside the sensor bore hole. 4. Allow 5-10 minutes for temperature to equilibrate.

Page 60 - Figure V.A.1

59 4. TRANSFER STANDARD METHODFor maximum accuracy of any Monitor/controller, the transfer standard method should be utilized. Instead of removin

Page 61

60VI. OPTIONS & ACCESSORIESA. OPTIONS ORDERED WITH MONITOR/CONTROLLERPART # DESCRIPTION -SC Second Alarm/control Module (Monitor/controllers

Page 62 - OEM INSTALLATIONS ONLY

61C. STANDARD SOLUTIONS & BUFFERS 1. CONDUCTIVITY/TDS STANDARD SOLUTIONSYour Conductivity/TDS Monitor/controller has been factory c

Page 63

62VII. REPLACEMENT PARTSMODEL DESCRIPTIONELECTRONIC MODULE CONDUCTIVITY756CB...756II Circuit Board Analog Monitor only7

Page 64 - VI. OPTIONS & ACCESSORIES

63VII. REPLACEMENT PARTS ContinuedENCLOSURE700II ECP ...750II Rear Housing (Box) for Picture Frame Front Panel700II ECR ...

Page 65 - DESCRIPTION

64VIII. WARRANTYAll Myron L Company Conductivity/TDS and Resistivity Monitor/controllers have a TWO Year Limited Warranty. If any Monitor/controller

Page 66 - VII. REPLACEMENT PARTS

65IX. GLOSSARY442™ An Internationally recognized “natural water” standard developed by the Myron L Company in 1964. Algorithm A procedure for solvi

Page 67 - MODEL DESCRIPTION

TABLE OF CONTENTS ContinuedSECTION PAGE 2. Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 69 - IX. GLOSSARY

67XI. ADDENDUMA. CONDUCTIVITY, TDS AND TEMPERATURE RELATIONSHIPS1. TEMPERATURE COMPENSATION (Tempco) of Aqueous SolutionsElectr

Page 70 - X. NOTES

68XI. ADDENDUM Continued

Page 71 - XI. ADDENDUM

69XI. ADDENDUM Continued

Page 72 - XI. ADDENDUM Continued

70XI. ADDENDUM Continued

Page 74

Water Quality InstrumentationAccuracy • Reliability • SimplicityMYRON L COMPANY2450 Impala DriveCarlsbad, CA 92010-7226 USATel: +1-760-438-2021Fax: +

Page 75

TABLE OF CONTENTS 5. Optional Front Panel Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 76 - Water Quality Instrumentation

Thank you for selecting one of the Myron L Company’s 750 Series II Monitor/controllers. The 750 Series II is based on input from ‘you’ - our customer