INSTALLATION • OPERATION • MAINTENANCEConductivity/TDS Models: 756II, 757II, 758II & 759IIResistivity Models: 751II, 752II, 753II & 754II750
NOTE: 3 1/2 digit LCD may be replaced with either a 3 1/2 or a 4 1/2 digit backlit LCD - available as options. ACCURACYAnalog Models: + 2% of Full Sca
E. SENSORS 1. CONDUCTIVITY/ TDS 750 Series II Conductivity/TDS Monitor/controllers use the CS51LC, CS51 or CS52 Series sensor depending on range.
RANGEResistivity0-20MΩ0-10MΩ†0-5MΩ†0-2MΩ0-1MΩ†0-500KΩ†0-200KΩConductivity0-1 µS*†0-1 ppm*†0-2 µS*†0-2 ppm*†0-5 µS*†0-5 ppm*†0-10 µS†0-10 ppm†0-
A. GENERALThis section provides the recommended procedures for properly installing the 750 Series II Conductivity/Resistivity Monitor/controller, a
4. Slide the enclosure through the panel cutout untilitsangecontactsthepanel. 5. Insertmountingscrewsthroughtheange mounting
D. SENSOR INSERTION/IMMERSION MOUNTING The Sensor’s mounting orientation must provide a continuous andadequatecirculationowtopreventthetrap
4. Turn the front panel around so that the back side is facing you and set aside for now. 5. Carefully remove front panel, leaving the harness
d. Install5.49kΩresistoratRED(RD)andthe NEUTRAL (NU) connector locations, as shown in gureII.E.5. 5. Carefullyreinstal
14 c. Connect to your own alarm or ?Use the following as guidelines. Connector is a standard 2 wire Methode* style connector. Connecto
F. 0-10 VDC RECORDER OUTPUTThe0-10VDCoutputisdesignedtogivetheuserthecapabilityof sending a signal to a remote meter, recorder, PLC or SC
i
16G. RE-RANGE YOUR MONITOR/CONTROLLER (Range Module Installation) 1. DESCRIPTIONThe 750 Series II Monitor/controllers have been design
17 3. CHANGING ANALOG METER SCALE (DIAL)After changing ranges on an analog model 750II it may be nec-essary to change the meter scale (dial). The
18 4. LABEL CHANGESince you have re-ranged your Monitor/controller it may be necessary to relabel it. If you made any one of the following changes
19A. SC/SCO MODULE (SECOND ALARM/CONTROL OPTION) (Digital & OEM Models ONLY)-SC Second Alarm Control Module ordered with
20Front PanelWARNING: There are two (2) O-rings installed on the switches, one (1) on the shank and the other is under the push button. Both of these
The following instructions describe the steps for converting the Monitor/controller.RefertogureIII.A.4forthelocationoftheSPC(s) on your Mon
c. Hysteresis (Dead Band) Adjustment*Thehysteresisordeadbandisapproximately±3%ofthesetpoint at full scale as it leaves the factory. Unde
B. 4A/4AO MODULE (4-20mA OPTION)Available only on the digital models 758II, 759II, 753II and 754II, or all analog and digital OEM Monitor/controlle
24FATAL TO PERSONNEL. ONLY QUALIFIED PERSON-NEL SHOULD INSTALL ELECTRICAL EQUIPMENT.PhysicalIf the front panel has all ready been removed from
25 5. If not, adjust the CAL control marked “4mA” until theDVMindicates4mA,seegureIII.B.3. 6. PresstheFullScaleTestSwitch,theD
If you read nothing else in this manual please read this Quick Reference Guide. PLEASE READ and COMPREHEND ALL WARNINGS, CAUTIONS and ADVISEMENTS CONT
4. CONVERTING A CURRENT TO A VOLTAGECurrent measuring devices actually measure voltage, but have aninternalresistorasshowni
27C. TEMPERATURE MODULE (Digital Models only)-TP Temperature Module ordered with Monitor/controller.TPO Temperature Module Kit ordere
28 5. Set front panel down or carefully allow to hang from the harness. Do not drop as the harness connector will pull out allowing the fr
29 a. TPC “Calibration” Module ProcedureThis procedure assumes the front panel is removed. 1. Ensure power is OFF. 2. Remove the sensor lead
30 19. Continue or reinstall the front panel and tightly secure four (4) retaining screws, see REASSEMBLY below.*CAUTION:Thesensorinputand0
31 counterclockwise position. If not, turn SCO Set Point control fully CCW - it may be 30 turns - or until it clicks. Forlocationseeg
32 5. OEM INSTALLATIONa. Single Display ModelsAn OEM installation requires an additional display harness, part number TPMH. This harness is insta
33 6. OEM INSTALLATIONb. Dual Display Models SPECIAL ORDER These models allow simultaneous display of both Cond/TDS, RES or pH/ORP and Temperatur
34D. ALARM/CONTROL KIT (Digital & OEM Models only) -TH Controller wiring Harness (requires second alarm/control module) orde
by pulling lightly on the wires. It may be necessary touseasmallatscrewdrivertoslightlyspreadthe retainingring.Seegu
connected to an appropriate Sensor, and that the Sensor is immersed in water within the range that the Monitor/controller will be required to read; an
c. Remove harness from SCO Module by pulling on harness,seegureIII.D.3. d. Set SCH harness aside, it is no longer used. 4. Re
8. Turn SCO Set Point control fully counter clockwise —itmaybe30turns-untilitclicks.SeegureIII.D.7 for location. 9. Locate
38Set Point Conversion (SPC) / Reversing Set PointTheMyronLCompanyshipstheTemperaturecontrollercong-ured to Alarm/control when the temperatur
39E. INSTALLATION OF BOTH TEMPERATURE MODULE & 4-20mA OUTPUT MODULE (Digital and OEM Models ONLY) -TP/TPO Temperature Modu
40F. 3 SENSOR INPUT MODULE3SO 3 Sensor Input Module with 3 foot / 1 meter cable ordered separately (customer installed and mounted).3SE
Module to the main Monitor/controller CB at the sensor input location as shown on label. 5. Attach the separate small cable harness from
42 gureIII.F.4.To disable the Sensor alarm/control function completely. 1. Locate the 8 position select switch on the 3S Module. 2. Set all
43G. 3 RANGE SWITCH MODULE 3RO 3 Range Switch Module with 3 foot / 1 meter cable ordered separately (customer installed and mounted).3RE
443ROModuleasanaccessoryyoumayneedtorecongurethe select switch for it to Range properly. 1. Locate the 8 position select switch on the
45H. 3 SENSOR / 3 RANGE MODULE 3SRO 3 Sensor / 3 Range Switch Module with 3 foot / 1 meter cable ordered separately (customer installed and mou
1750 Series IIModel 758 -121-SC(A Digital ConductivityMonitor/controller,with a Range of 0-2000 µS,and a Second Alarm/control,Displaying Full Scale
46Sensor ConnectionsFor convenience, it is recommended you connect the Primary or controlling sensor to SENSOR 1 position as shown in gureIII.H.1.
47For Conductivity/TDS - X1, X10 and X100. For Resistivity - X1, 0.1, and 0.01. If adding an alarm/control range. 1. Locate the 8 position select s
48I. PIEZO ELECTRIC ALARM -PA Piezo Electric Alarm ordered with Digital Monitor/controller.PAO Piezo Electric Alarm ordered separate
49 a. For a Conductivity/TDS Monitor/controller, as shipped from the Myron L Company, the alarm will sound off when the Full Scale Te
50J. REMOTE ALARM - RA™ RA Ordered as an accessory, includes 8” harness with connector, and two (2) wire nuts. 1. DESCRIPTIONThisremote
51 alarm for approximately three minutes or until you improve the water quality (readjust controller set point). Ifthreeminutesmuting
52IV. OPERATING PROCEDURESA. FRONT PANEL INDICATORS & CONTROLSThe front panel illustrations to the right, switch and indicator operational desc
53B. OEM FRONT PANEL INDICATORS & CONTROLSThe simulated OEM front panel illustrations to the right, and switch and indicator
54C. SETUP PROCEDURES These Setup procedures cover (1) converting the alarm circuit to trigger on a decreasing (conductivity) or increasing (Resist
55CAUTION: adjusting the hysteresis too tight may cause the alarmtouctuate(on-off)duetoow,chemicalmixingorbub-bles causing the relay to ch
TABLE OF CONTENTS 750 Series II ILLUSTRATION (758II-121-SC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
56A. PRIMARY COMPONENT IDENTIFICATIONAsidentiedinSectionIII,theMonitors/controllersswitch(es)and indicator components are mounted directl
57B. METER MECHANICAL ZERO PROCEDURESANALOG METERS ONLY 1. Ensure power is OFF. 2. Note the position of the meter needle. If it has come to
58 3. Slowly shake the sensor to remove air bubbles from inside the sensor bore hole. 4. Allow 5-10 minutes for temperature to equilibrate.
59 4. TRANSFER STANDARD METHODFor maximum accuracy of any Monitor/controller, the transfer standard method should be utilized. Instead of removin
60VI. OPTIONS & ACCESSORIESA. OPTIONS ORDERED WITH MONITOR/CONTROLLERPART # DESCRIPTION -SC Second Alarm/control Module (Monitor/controllers
61C. STANDARD SOLUTIONS & BUFFERS 1. CONDUCTIVITY/TDS STANDARD SOLUTIONSYour Conductivity/TDS Monitor/controller has been factory c
62VII. REPLACEMENT PARTSMODEL DESCRIPTIONELECTRONIC MODULE CONDUCTIVITY756CB...756II Circuit Board Analog Monitor only7
63VII. REPLACEMENT PARTS ContinuedENCLOSURE700II ECP ...750II Rear Housing (Box) for Picture Frame Front Panel700II ECR ...
64VIII. WARRANTYAll Myron L Company Conductivity/TDS and Resistivity Monitor/controllers have a TWO Year Limited Warranty. If any Monitor/controller
65IX. GLOSSARY442™ An Internationally recognized “natural water” standard developed by the Myron L Company in 1964. Algorithm A procedure for solvi
TABLE OF CONTENTS ContinuedSECTION PAGE 2. Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . .
66X. NOTES
67XI. ADDENDUMA. CONDUCTIVITY, TDS AND TEMPERATURE RELATIONSHIPS1. TEMPERATURE COMPENSATION (Tempco) of Aqueous SolutionsElectr
68XI. ADDENDUM Continued
69XI. ADDENDUM Continued
70XI. ADDENDUM Continued