AQUASWITCHAQUASWITCH & 760-A AQUASWITCHOperation ManualINSTALLATION • OPERATION • MAINTENANCE22 August 05Water Quality InstrumentationAccuracy • R
SECTION 3Operating Procedures3.1 GENERALSection 3.2 provides the operational descriptions for each of the Aquaswitch/760-A Aquaswitch switch a
3.3 OPERATIONAL CHECKOUT PROCEDURESThe following checkout procedures are used to verify that the Aquaswitch or 760-A Aquaswitch is operating
SECTION 4Component Identication,Calibration and Preventive Maintenance4.1 GENERALSection 4.2 provides detailed illustrations to assist the us
4.3 “PURGE CYCLE”. CALIBRATION PROCEDURESNOTE:The Purge Cycle trimmer adjustment screw (R2) (see Fig. 4-1) determines the time span that the
4.4 “ALARM RESET” JUMPER ADJUSTMENTNOTE:The Aquaswitch/760-A Aquaswitch are shipped with an Alarm condition that occurs whenever the “IN USE” Bank
4.5.3 FULL SCALE CALIBRATION PROCEDURESNOTE:Before attempting to calibrate the 760-A Aquaswitch analog meter, refer to Section 4.5.1.STEP 1 Turn OFF
WARRANTYAll Myron L monitors and cells have a one year warranty. If any monitor or cell fails to function normally, return the faulty unit to the fact
IMPORTANT! PLEASE READ!ADDENDUM TO USER’S MANUALAQUASWITCH 760A AQUASWITCHMost Aquaswitch users choose Myron L HP Series 24 vac valves. For this re
Water Quality InstrumentationAccuracy • Reliability • SimplicityACCURACY • RELIABILITY • SIMPLICITYMYRON L COMPANY2450 Impala DriveCarlsbad, CA 92010-
SECTION PAGE1
SECTION 1Introduction1.1 SCOPEThis manual provides the necessary information to install, operate and maintain the Myron L Company’s Aquaswi
1.2.3 HP SOLENOID PILOT VALVESDESCRIPTION: Being corrosive, ultra-pure water can be difcult to control. Myron L HP valves solve this problem by utili
SECTION 2Installation2.1 GENERALSection 2.2 provides the user with the recommended procedures for mounting the Aquaswitch/760-A Aquaswitch and for com
NOTE:The 760-A Aquaswitch is not designed to operate with a Cell cable length that exceeds 100' (30 meters).3 If at all possible, mount
2.3 ELECTRICAL INSTALLATIONThe Aquaswitch and 760-A Aquaswitch require the user to follow separate electrical installation procedures. In addition, ba
2.3.2 760-A AQUASWITCH MAIN AC POWERNOTE:The following procedures are to be used to install the 760-A Aquaswitch 115 VAC main power source. For the pr
2.3.4 PILOT VALVE SOLENOID CONNECTIONSNOTE:Section 2.3.4 provides the procedures for installing 115 VAC solenoid valves with maximum load rates of 25